The Tribioscience Vero cell HCP Fast ELISA is designed to quantitatively detect Vero cell HCP residues in different biological samples. The main feature is that the kit uses our novel proprietary approaches to combine samples and detections into a one-step instead of the complicated traditional methods. It makes the assay flexible, accurate, simple, and time saving. The measurement can be finished in 2 hours, not needing 4-5 hours (Fig. 1). The detection range is from 1 to 200 ng/mL. The levels of Vero HCP samples are parallel to the standard curves obtained using the kit standards linearly. These results indicate that this kit can be used to determine relative mass values for Vero host cell protein.
Typical Data
This standard curve is provided for demonstration only. A standard curve should be generated for each set of samples assayed. Fig. 3 is an example of typical Data.
Sensitivity
LOD is about 1.1 ng/mL
LOQ is 4.9 ng/mL
The Intra-assay CV is 3.79% and the Inter-assay CV is ˂10%.
Specificity
This assay recognizes Vero HCP.
Simple Procedures
Simple Standard Preparation
Standard Curve
Kit Content and Storage Conditions
PART
PART#
DESCRIPTION
STORAGE OF OPENED/ RECONSTITUTED
Vero HCP Microplate
TBS31005A
96 well polystyrene microplate (12 strips of 8 wells) coated with a polyclonal antibody specific to Vero HCP.
Return unused wells to the foil pouch. Reseal along the entire edge of the zip-seal. May be stored for up to 1 month at 2-8 °C.
Vero HCP Standard
TBS31005B
90 µL of Vero HCP protein (2 µg/mL).
Aliquot and store at -20 ℃ for up to 1 month in a manual defrost the freezer. Avoid repeated freeze-thaw cycles.
Detection A
TBS31005C
55 µL of HRP Vero HCP antibody.
May be stored for up to
3 months at 2-8 °C.*
Assay Diluent
TBS31005D
25 mL of a buffered protein base with preservatives.
Wash Buffer
TBS3000W
15 mL of concentrated solution (10x).
TMB Substrate
TBS3000T
12 mL of ultra-sensitive TMB substrate.
Stop Solution
TBS3000S
6 mL of 2 N sulfuric acid.
Store the unopened kit at 2-8 ℃. Do not use past kit expiration date.
The kit contains sufficient materials to run an ELISA on one 96 well plate.